The implant's map sets, for each electrode, the softest current that is heard (T) and the loudest current that is still comfortable (C or M), and which frequency band it receives. The external processor splits sound into bands and transmits through the skin, by radio, to the internal implant that stimulates the electrodes in the scala tympani.

Why it matters in the exam

Components: outside — microphone, speech processor and transmitting coil; under the skin — a receiver-stimulator with antenna and magnet, and the electrode array, which the surgeon threads into the scala tympani via the round window or a small cochleostomy. Signal and power cross the skin over a radio-frequency (RF) link, and telemetry is the implant reporting back to the processor. The T level is the softest current the user detects — set with a detection task; the C/M level is the loudest current that is still comfortable, set with loudness scaling. The electrical dynamic range — from T to C/M — is very narrow, roughly 10–20 dB. Mapping starts about two to four weeks after surgery, when the wound has healed and the patient is awake and cooperative.

The coding strategy — ACE, CIS or FSP, for example — is how the processor analyses sound and turns it into pulses. Electrodes follow tonotopy: basal ones receive high frequencies and apical ones low; a user who cannot hear /s/ needs a check of the basal electrodes that carry the highs. If the array was not fully inserted, the apical electrodes are missing and the lows suffer. To get the total stimulation rate, multiply the active channels by each channel's rate: 12 electrodes at 1200 each give 14,400 pulses every second. Facial twitching with one electrode — current spreading to the facial nerve: lower the level or switch the electrode off. Objective measures for infants: the ECAP (the nerve response recorded through the implant's own electrodes) helps the first map, and the eSRT (stapedius reflex threshold to electrical stimulation) relates closely to C/M levels. Impedance testing shows whether each electrode is sound — open or short circuit — but does not set T or C. A user who complains of noise — start with a listening check of the external microphone.

Key points

Summary table

ConceptDefinitionUse
T levelThe softest current detectedA detection task
C/M levelThe loudest current still comfortableLoudness scaling; the eSRT guides it
Electrical dynamic rangeFrom T to C/M, roughly 10–20 dBFar narrower than a hearing ear
Frequency allocationBase — highs; apex — lowsTrouble with /s/ — check basal electrodes
Impedance and telemetryElectrode integrity; implant reports to processorDo not set T or C
ECAP and eSRTNerve response through the implant; electrical stapedius reflexA map for an infant who cannot respond
Stimulation rateActive channels multiplied by per-channel rate12 electrodes at 1200 → 14,400

In the full version for subscribers

The full version is open to subscribers of: Speech and hearing (communication disorders).

Subscriber? Read the full topic Plans and prices 25 free questions

Sources

Sources are given for further reading; the text is written afresh and not copied from them.

Reviewed by the Examqo AI Professor · 2026-11-02. The text may contain errors; when in doubt, check the source.