Otoacoustic emissions (OAE): TEOAE, DPOAE and SOAE

OAEs are faint sounds the cochlea itself produces thanks to outer hair cell motility, and they travel back through the middle ear to a microphone in the canal. A normal response says the middle ear and the outer cells work — not that the nerve is normal, and not what the threshold is.

Why it matters in the exam

The bank asks what is needed to obtain emissions: a normal middle ear in both directions and active outer cells. So present emissions rule out significant middle-ear fluid and a moderate or greater cochlear loss in that region, but not auditory neuropathy — the false negative of screening. Absent emissions may come from fluid, cochlear loss or a blocked canal, and do not show the degree of loss. OAEs are unaffected by sleep and sedation, so they suit infants; they are not a threshold test. Deciding that a response is present rests on a signal-to-noise ratio of at least about 6 dB in several frequency bands, together with good reproducibility.

How the types differ: TEOAEs are evoked by a broadband click, give information mainly at 1000–4000 Hz and usually vanish once the loss passes about 20–30 dB HL. DPOAEs are evoked by two tones (f2 ÷ f1 roughly 1.2); in humans the largest emission is at 2f1 − f2 (f1 = 2000, f2 = 2400 → 1600 Hz), they can be measured up to about 8000–10,000 Hz, and they sometimes persist with slightly greater losses (up to about 40 dB HL) — so DPOAEs are sometimes present without TEOAEs. SOAEs occur with no stimulus in a sizeable share of normal ears, so their absence is not pathological. Uses: newborn screening, ototoxic and noise monitoring (an early change before the audiogram), suspected functional loss, diagnosing ANSD and testing the efferent system.

Key points

Summary table

TypeStimulus and featuresUse and limit
TEOAEBroadband click; information mainly 1000–4000 HzVanish above about 20–30 dB HL; newborn screening
DPOAETwo tones, f2 ÷ f1 roughly 1.2; emission at 2f1 − f2Up to about 8000–10,000 Hz; ototoxic monitoring
SOAENo stimulus; narrow tonesMissing in many normal ears too — not diagnostic
Response criterionSignal-to-noise ratio of at least 6 dB and good reproducibilityIn several frequency bands, not just one
Present emissionsOuter cells and middle ear workingDo not rule out ANSD; do not set a threshold
Absent emissionsFluid, cochlear loss or blockageDo not show the degree of loss

In the full version for subscribers

The full version is open to subscribers of: Speech and hearing (communication disorders).

Subscriber? Read the full topic Plans and prices 25 free questions

Sources

Sources are given for further reading; the text is written afresh and not copied from them.

Reviewed by the Examqo AI Professor · 2026-11-02. The text may contain errors; when in doubt, check the source.