PET and PET/CT physics

A positron travels a short distance, annihilates with an electron and creates two 511 keV photons in nearly opposite directions; the scanner detects them simultaneously (coincidence) and draws a line between them — "electronic collimation" that gives far higher sensitivity than SPECT.

Why it matters in the exam

The exam asks about event types: true — both photons from the same event, without scatter; scatter — one photon was deflected and the line is misplaced; and random — two photons from different events detected by chance in the same timing window. Random events rise roughly with the square of activity and are corrected with a delayed window. Factors that limit resolution: positron range (a higher-energy positron travels further — Rb-82 and Ga-68 are less sharp than F-18), slight non-collinearity of the two photons, and detector size.

Detectors: LSO or LYSO crystals are fast and bright, and BGO is dense; in "digital" PET the PMTs have been replaced by silicon photomultipliers (SiPM). Time of flight measures the difference in arrival time of the two photons, narrows the event position on the line and improves the signal-to-noise ratio, mainly in large patients. Attenuation correction is CT-based; so metal, concentrated contrast and respiratory motion (misregistration at the lung base and liver dome) create artefacts — and the non-corrected image should always be reviewed too.

SUV = tissue activity concentration divided by (injected activity divided by body weight); it depends on calibration, time from injection, glucose and lesion size (partial volume effect in lesions smaller than about twice the resolution). CT radiation dose is measured by CTDIvol (average dose in a slice) and by DLP, which is CTDIvol multiplied by the scan length and therefore accounts for length; effective dose is estimated from it. Low-dose CT for attenuation correction only reduces radiation when diagnostic CT is not needed. Daily quality control includes a detector check and calibration with a source or phantom.

Key points

Summary table

ConceptDefinitionEffect
True eventTwo photons from the same eventThe wanted signal
Scatter eventA photon was deflectedWrong position, lower contrast
Random eventPhotons from different eventsRises roughly with the square of activity
Positron rangeDistance to annihilationHigh energy — poorer resolution
Time of flightDifference in arrival timesBetter signal-to-noise ratio
DLPCTDIvol multiplied by scan lengthAccounts for length

A free exam question

במדידת Hounsfield units של ממצא ב- CT מתקבל ערך של אפס . מהו היחס בין מקדם הבליעה של הקרינה של מים ( m ) לזו של הממצא ( =? m / m )? water water (the question as examined, in Hebrew)

  1. א. 0
  2. ב. 1
  3. ג. 2
  4. ד. 1000

See the answer and explanation

In the full version for subscribers

The full version is open to subscribers of: Nuclear medicine, Diagnostic radiology.

Subscriber? Read the full topic Plans and prices 25 free questions

Sources

Sources are given for further reading; the text is written afresh and not copied from them.

Reviewed by the Examqo AI Professor · 2026-11-06. The text may contain errors; when in doubt, check the source.