A gamma camera converts a gamma photon into light in an NaI(Tl) crystal, light into an electrical signal in photomultiplier tubes (PMT), and accepts only photons of the right energy (PHA) — and the collimator, which accepts only photons travelling in one direction, sets the balance between resolution and sensitivity.

Why it matters in the exam

The exam asks about each component's role. The parallel-hole collimator defines direction and without it there is no image; its resolution is better with longer, narrower holes, but sensitivity then falls; and the most important factor is distance from the patient — the closer the camera, the sharper the image. High-energy isotopes (I-131, In-111, Ga-67) need a collimator with thicker septa. A pinhole collimator magnifies small organs such as the thyroid; a converging (fan beam) collimator improves brain SPECT.

The PHA (pulse-height analyser) accepts only signals within an energy window (for example 20% around 140 keV) and rejects scattered photons that lost energy. A semiconductor detector (CZT) converts a photon directly into an electrical signal, without a light stage, so its energy resolution is better; in dedicated cardiac cameras the high sensitivity also comes from geometry focused on the heart. SPECT: the camera rotates around the patient and collects projections — usually 60–128 over 360°, and for the heart usually 180°. A body-contour orbit keeps the distance short and improves resolution compared with a circular orbit.

Reconstruction: filtered back-projection (FBP) is fast; iterative reconstruction (OSEM) is more accurate and reduces artefacts, but entered routine use late because of computation time. A low-pass filter smooths noise; a high-pass filter enhances edges and also noise. SPECT/CT adds anatomical localisation and attenuation correction. Quality control: daily uniformity — non-uniformity causes ring artefacts in SPECT; resolution and linearity with a bar phantom weekly; centre of rotation for SPECT; and daily energy peaking. A failed PMT appears as a round cold spot.

Key points

Summary table

ComponentRoleKey point
Parallel collimatorSelecting directionLong and narrow — good resolution, low sensitivity
NaI(Tl) crystalConverting gamma to lightThick crystal — high sensitivity, lower resolution
PMTConverting light to an electrical signalFailed PMT — cold spot
PHASelecting energyRejects scattered photons
CZTDirect conversion to an electrical signalGood energy resolution
CT in a hybrid systemLocalisation and attenuation correctionMetal and motion artefact

In the full version for subscribers

The full version is open to subscribers of: Nuclear medicine, Diagnostic radiology.

Subscriber? Read the full topic Plans and prices 25 free questions

Sources

Sources are given for further reading; the text is written afresh and not copied from them.

Reviewed by the Examqo AI Professor · 2026-11-06. The text may contain errors; when in doubt, check the source.

See also: Basics of diagnostic nuclear medicine