CT principles: HU values, window, pitch, noise and artefacts

In CT every voxel receives a value in Hounsfield units (water 0, air -1000), the window decides what is seen, pitch decides speed and dose, noise falls with mAs and thicker slices — and artefacts come from beam hardening, metal, motion or a faulty detector.

Why it matters in the exam

CT questions in the exam are short and technical. HU values: water 0, air -1000, fat about -100 to -50, soft tissue 20–60, fresh blood about 50–80, and bone above 400. Window width sets the HU range displayed in grey levels: a narrow window increases contrast (brain, liver), a wide window lowers contrast and shows a large range (lung, bone). The window level should be near the HU of the tissue examined. Pitch is table travel per rotation divided by total beam width; a pitch above 1 shortens the scan and lowers dose, at the cost of resolution along the axis.

Quantum noise depends on the number of photons reaching the detector: it falls when mAs or kV is raised, with thicker slices and larger voxels, and with a soft reconstruction kernel; it rises in a large patient. Fine detail is seen with thin slices, a sharp (bone) kernel, a small FOV and a small focal spot. Iterative reconstruction reduces noise compared with filtered back projection, and so allows a lower dose. Dual-energy CT separates materials: iodine maps, "virtual non-contrast" images, distinguishing uric acid from calcium stones, virtual monoenergetic images and reduction of metal artefact.

Artefacts: beam hardening — dark streaks between two dense structures (for example between the petrous bones at the skull base); metal artefact — bright and dark streaks around implants; motion — doubling and blurring; partial volume — an averaged value between two tissues in a thick slice; ring artefact — a miscalibrated or faulty detector. The "bow-tie" filter lowers dose at the edges of the body and balances the signal to the detectors. The anode–cathode axis lies along the Z axis so the heel effect does not alter the beam across the patient. Dose metrics: CTDIvol in mGy and DLP in mGy·cm. Injection phases: arterial about 25–35 seconds, portal venous about 60–70 seconds, delayed 3–5 minutes.

Key points

Summary table

TopicWhat it determinesChange that improves itCost
Window widthDisplayed contrastNarrow window → high contrastA small HU range is seen
PitchSpeed and doseHigh pitch → fast, low doseResolution along the axis falls
NoiseNumber of photonsHigh mAs, thick slice, iterative reconstructionDose or sharpness
Spatial resolutionFine detailThin slice, sharp kernel, small focal spotMore noise
ArtefactsImage qualityRecognise: hardening, metal, motion, ringRing = faulty detector

In the full version for subscribers

The full version is open to subscribers of: Diagnostic radiology, Medical licensing exam (IMLE).

Subscriber? Read the full topic Plans and prices 25 free questions

Sources

Sources are given for further reading; the text is written afresh and not copied from them.

Reviewed by the Examqo AI Professor · 2026-11-06. The text may contain errors; when in doubt, check the source.