X-ray image formation: tube, beam and image quality

In the tube X-rays arise mainly from braking of electrons (bremsstrahlung), kV sets beam energy and contrast, mAs sets the number of photons and the noise — and the filter, grid and geometry set dose, contrast and sharpness.

Why it matters in the exam

X-ray physics questions repeat the same components. In the tube, electrons leave the heated cathode and strike the anode (usually tungsten). Most radiation is bremsstrahlung with a continuous spectrum; characteristic radiation appears at fixed energies set by the anode material — this is the answer when asked "what determines characteristic radiation". The heel effect: beam intensity is lower on the anode side, so the thicker body part is placed on the cathode side. Mammography uses a molybdenum, rhodium or tungsten anode and filter at 25–32 kV to obtain low energies that increase soft-tissue contrast.

The two important interaction processes: the photoelectric effect — the photon is fully absorbed and releases an inner-shell electron; its probability rises steeply with atomic number (about Z³) and falls with energy (about 1/E³), and it is the source of contrast between bone, iodine and tissue. Compton scatter — the photon loses part of its energy and changes direction; it is the commonest interaction in soft tissue at radiography and CT energies, lowers contrast and is the main source of staff dose. The (aluminium) filter removes low-energy photons that the skin would absorb without adding anything to the image — it reduces skin dose and "hardens" the beam.

The grid absorbs scattered radiation and improves contrast, but requires a higher dose; it can be omitted for thin body parts (limbs, small children). A focused grid whose distance does not match the focal distance causes grid cut-off. Raising kV lowers contrast and skin dose; raising mAs lowers noise but does not improve contrast. Increasing the object–detector distance increases magnification and blur (except for the air-gap technique, which reduces scatter); good sharpness comes from a small focal spot, a short object–detector distance and a long focus–detector distance. In CR the latent image is stored in a phosphor plate and read by a laser; in DR a flat-panel detector reads directly, or through a scintillator layer that converts X-rays into light. With automatic exposure control (photo-timer), too low a kV lengthens the exposure and gives the patient a higher dose.

Key points

Summary table

Parameter or componentWhat it changesDirection of changeTypical question
kVBeam energy and penetrationHigh kV → low contrast, low skin doseWhy mammography uses low kV
mAsNumber of photonsHigh mAs → less noise, higher doseDoes mAs improve contrast? No
FilterRemoves soft photonsLow skin dose, "hard" beamRole of the filter in the tube
GridAbsorbs scattered radiationGood contrast, higher doseWhen it is omitted
Focal spot and distancesSharpness and magnificationSmall focus and short object–detector distance → sharpWhat happens when object–detector distance increases

Free exam questions

מה תפקיד הסריג( GRID ) במערכת DR ? (the question as examined, in Hebrew)

  1. א. מקצר את זמן הצילום ומונע תזוזה
  2. ב. גורם לקבלת תמונה בעלת ניגודיות טובה
  3. ג. מקטין את חשיפת החולה לקרינה
  4. ד. מעלה את כשר ההפרדה המרחבי של התמונה

See the answer and explanation

מי מהשיטות הבאות לא משפרת קונטרסט ברקמה? (the question as examined, in Hebrew)

  1. א. הורדת kV
  2. ב. העלאת mA
  3. ג. שימוש בסורג( grid .)
  4. ד. מתן חומר ניגוד בוריד

See the answer and explanation

In the full version for subscribers

The full version is open to subscribers of: Diagnostic radiology, Medical licensing exam (IMLE).

Subscriber? Read the full topic Plans and prices 25 free questions

Sources

Sources are given for further reading; the text is written afresh and not copied from them.

Reviewed by the Examqo AI Professor · 2026-11-06. The text may contain errors; when in doubt, check the source.