Monitoring in anaesthesia — oximetry, capnography and depth of anaesthesia

Basic monitoring in every anaesthetic covers oxygenation, ventilation, circulation and temperature — and whoever can read the capnography trace and knows the limits of pulse oximetry catches most mishaps before they become disasters.

Why it matters in the exam

Questions show a capnography waveform or a sudden change in EtCO2 and ask for the cause: disconnection, oesophageal intubation, embolism, bronchospasm, rebreathing or malignant hyperthermia. Others test when pulse oximetry is wrong (CO poisoning, methaemoglobinaemia, dyes), the target BIS value, and what happens to the arterial-line pressure when the transducer is at the wrong height.

Under the ASA standards, every general anaesthetic measures inspired oxygen, pulse oximetry, continuous capnography, a continuous ECG and blood pressure at least every 5 minutes, and temperature when changes are expected. EtCO2 is usually 2–5 mmHg below PaCO2; the gap widens when dead space grows — in pulmonary embolism, low cardiac output and lung disease. A saturation of 90% corresponds roughly to a PaO2 of 60 mmHg, and below that point the curve is steep.

In exam questions remember the order in time: capnography responds at once to a ventilation problem, while pulse oximetry responds late, especially after preoxygenation. So a disconnection or an oesophageal tube is detected first on capnography. Consciousness is monitored too: awareness under anaesthesia is more common with intravenous anaesthesia plus paralysis, at caesarean section and in a hypotensive trauma patient, and depth-of-anaesthesia monitoring helps in these settings.

Key points

Summary table

FindingLikely meaningExam point
EtCO2 drops to zero at onceDisconnection, oesophageal intubation, complete obstruction, cardiac arrest"No trace = tube not in place" until proven otherwise
EtCO2 drops sharply but not to zeroPulmonary or air embolism, sudden fall in cardiac outputThe PaCO2–EtCO2 gap widens
EtCO2 rises graduallyHypoventilation, CO2 absorption in laparoscopy, fever, malignant hyperthermiaAn unexplained rise with tachycardia → malignant hyperthermia
Baseline does not return to zeroRebreathing: exhausted CO2 absorbent or a faulty valveChange the absorbent, check the valves
Shark-fin waveformBronchospasm or obstruction to expiratory flowUpsloping phase 3

In the full version for subscribers

The full version is open to subscribers of: Anesthesiology, Emergency medicine, Medical licensing exam (IMLE).

Subscriber? Read the full topic Plans and prices 25 free questions

Sources

Sources are given for further reading; the text is written afresh and not copied from them.

Reviewed by the Examqo AI Professor · 2026-11-02. The text may contain errors; when in doubt, check the source.