The anaesthesia machine, gases and operating room fire

The anaesthesia machine is built with layers of protection against a hypoxic mixture, but only the oxygen analyser in the inspiratory limb detects every failure — and next to it a self-inflating bag and an oxygen cylinder must always be at hand.

Why it matters in the exam

The exam asks about the machine's safety systems (pin index, hypoxic guard, fail-safe valve), the difference between an oxygen cylinder and an N2O cylinder (gas versus liquid), a failure in the oxygen pipeline and what to do, the CO2 absorbent and the signs that it is exhausted, Mapleson circuits, the heated vaporiser for desflurane, and airway fire — the order of actions.

In an oxygen cylinder the gas is compressed, so pressure falls in direct proportion to the amount left. N2O is held in the cylinder as a liquid, and the pressure stays constant until the liquid is used up — so the amount is known only by weighing. In the circle system there are one-way valves and a CO2 absorbent; when the absorbent is exhausted CO2 appears in the inspired gas and the capnography baseline does not return to zero. Dry absorbent with desflurane can produce CO, and with sevoflurane compound A forms and, with dry absorbent, heat as well.

An operating room fire needs three components: an oxidiser (oxygen or N2O), fuel (drapes, the tracheal tube, alcohol-based skin preparation) and an ignition source (electrosurgery or laser). In head and neck surgery the oxygen concentration is kept as low as possible, the skin preparation is allowed to dry, and for airway laser surgery a laser-resistant tube with a saline-filled cuff is used. In an airway fire the tube is removed, gases are stopped, saline is poured, and then the airway is re-established and checked by bronchoscopy.

Key points

Summary table

IssueMain pointExam point
Machine protectionsPin index for cylinders, unique pipeline connectors, fail-safe, hypoxic guardOnly the oxygen analyser detects a pipeline crossover
CylindersOxygen — gas, pressure reflects content; N2O — liquid, constant pressure until the endN2O content is known by weighing
Circuit and absorbentOne-way valves, CO2 absorbent changes colour when exhaustedInspired CO2 = exhausted absorbent or a faulty valve
VaporisersAgent-specific and temperature-compensated; desflurane — heated, pressurised vaporiserHigh vapour pressure and boiling point near room temperature
Airway fireRemove the tube, stop gases, saline, re-ventilate, bronchoscopyPrevention: low oxygen, dry skin preparation, laser tube

A free exam question

בבדיקת מכונה לפני כל מקרה הרדמה, מה המרדים מחויב לבדוק על פי המלצות ASA ? (the question as examined, in Hebrew)

  1. א. לבדוק שהלחץ בבלון החמצן הרזרבי גבוה מ- 1000 PSIG
  2. ב. לבדוק שהלחץ בצינורות הגזים המגיעים לחדר גבוה מ- 50 PSIG
  3. ג. לבדוק את הלחץ במערכת ההנשמה ומבחן דליפה
  4. ד. לעשות קליברציה לסנסור החמצן

See the answer and explanation

In the full version for subscribers

The full version is open to subscribers of: Anesthesiology.

Subscriber? Read the full topic Plans and prices 25 free questions

Sources

Sources are given for further reading; the text is written afresh and not copied from them.

Reviewed by the Examqo AI Professor · 2026-11-02. The text may contain errors; when in doubt, check the source.